In a world increasingly facing new challenges at the forefront of plasma scientific research and technological innovation, CNR and ISTP pledge progress and achieve an impact in the integration of research into societal practices and policy

Millimeterwave reflectivity tests on MetOp-SG MWI On Board Calibration Target

Simonetto A.; Cuttaia F.; Biggi M.; Lapini R.; Lupi T.; Mandolesi N.; Matticari G.;
Morgante G.; Salaris P.; Salghetti Drioli L.; Terenzi L.

he On-Board Calibration Target (OBCT) for the Micro-Wave Imager instrument on the MetOp-SG-B satellite is a microwave blackbody, operating between 14 and 220 GHz (extended frequency range), providing the MWI receivers with a warm calibration reference. It is designed for an emissivity better than 0,9999 over the MWI 26 channels (including dual polarization channels). The OBCT radiometric performance has been extensively verified in ISTP (formerly IFP) microwave laboratory, by measuring reflectivity over the extended bandwidth. Measurement techniques and results are described in this paper, showing that the OBCT models tested so far are fully compliant with the MWI requirements. © 2021 IOP Publishing Ltd and Sissa Medialab.

ID 452417
DOI 10.1088/1748-0221/16/03/P03036
PRODUCT TYPE Journal Article
LAST UPDATE 2022-04-11T15:48:30Z
TOP