In a world increasingly facing new challenges at the forefront of plasma scientific research and technological innovation, CNR and ISTP pledge progress and achieve an impact in the integration of research into societal practices and policy

Spark discharge-libs: Evaluation of one-point and multi-voltage calibration for p and al determination

Vieira A.L.; Ferreira E.C.; Junior D.S.; Senesi G.S.; Neto J.A.G.

Spark discharge (SD) laser-induced breakdown spectroscopy (LIBS) is a technique suitable to overcome the low energies of lasers by reheating the plasma, increasing the emission intensities and to perform single-standard calibration. A calibration method called one-point and multi-voltage calibration (OP-MVC), which requires two different voltages applied to both the standard and the sample, is proposed for use with SD-LIBS. The performance of this method was compared to that of the one-point and multi-lines calibration (OP-MLC) and the slope ratio calibration (SRC) methods for LIBS determination of Al in certified reference plant leaves and P in commercial fertilizers. No statistical differences at the 95% confidence level were observed between the Al and P concentrations determined by OP-MVC LIBS for the Al certified values and the P values measured by high-resolution continuum-source flame atomic absorption spectrometry (HR-CS FAAS). The limit of detection (LOD) for P was 0.60 wt% P2O5 and 35.1 mg kg(-1) for Al. The relative standard deviation (n=3) was typically 7% for Al and in the 4 – 10% range for P.

ID 456989
DOI 10.46770/AS.2020.202
PRODUCT TYPE Journal Article
LAST UPDATE 2022-04-11 16:09:42.000