In a world increasingly facing new challenges at the forefront of plasma scientific research and technological innovation, CNR and ISTP pledge progress and achieve an impact in the integration of research into societal practices and policy

Novel compact hard x-ray spectrometer with MCps counting rate capabilities for runaway electron measurements on DIII-D

Dal Molin A.; Fumagalli L.; Nocente M.; Rigamonti D.; Tardocchi M.; Giacomelli L.; Panontin E.; Lvovskiy A.; Paz-Soldan C.; Edietis N.W.; Gorini G.

A novel compact spectrometer optimized for the measurement of hard x rays generated by runaway electrons is presented. The detector is designed to be installed in the fan-shaped collimator of the gamma-ray imager diagnostic at the DIII-D tokamak. The spectrometer is based on a 1 × 1 cm2 cerium doped yttrium aluminum perovskite scintillator crystal coupled with a silicon photomultiplier. The detector dynamic energy range is in excess of 10 MeV, with an energy resolution of ~10% at 661.7 keV. The fast detector signal (?70 ns full width at half maximum) allows for operation at counting rates in excess of 1 MCps. The gain stability of the system can be monitored in real time using a light-emitting diode embedded in the instrument. The detector is expected to be deployed in the forthcoming DIII-D runaway electron experimental campaign.

ID 453592
DOI 10.1063/5.0043762
PRODUCT TYPE Journal Article
LAST UPDATE 2022-04-11T18:14:52Z
EU PROJECT EUROfusion
TITLE Implementation of activities described in the Roadmap to Fusion during Horizon 2020 through a Joint programme of the members of the EUROfusion consortium
FOUNDING PROGRAM H2020