Runaway electron gamma-ray detection system, a novel hard x-ray (HXR) spectrometer optimized for bremsstrahlung radiation measurement from runaway electrons in fusion plasmas, has been developed. The detector is based on a 1’×1′ LaBr3:Ce scintillator crystal coupled with a photomultiplier tube. The system has an energy dynamic range exceeding 20 MeV with an energy resolution of 3% at 661.7 keV. The detector gain is stable even under severe loads, with a gain shift that stays below 3% at HXR counting rates in excess of 1 MCps. The performance of the system enables unprecedented studies of the time-dependent runaway electron energy distribution function, as shown in recent runaway electron physics experiments at the ASDEX Upgrade and COMPASS tokamaks.
A new hard x-ray spectrometer for runaway electron measurements in tokamaks
Dal Molin A.; Nocente M.; Dalla Rosa M.; Panontin E.; Rigamonti D.; Tardocchi M.; Shevelev A.; Khilkevitch E.; Iliasova M.; Giacomelli L.; Gorini G., Perelli Cippo E.; D’Isa F.; Pautasso G., Papp G., Tardini G.; Macusova E., Cerovsky J.; Ficker O.; Salewski M.; Kiptily V.; EUROfusion MST1 Team; COMPASS Team; ASDEX Upgrade Team
| ID | 481629 |
|---|---|
| DOI | 10.1088/1361-6501/acd46c |
| PRODUCT TYPE | Journal Article |
| LAST UPDATE | 2023-06-19T10:03:20Z |
| EU PROJECT | EUROfusion |
|---|---|
| TITLE | Implementation of activities described in the Roadmap to Fusion during Horizon 2020 through a Joint programme of the members of the EUROfusion consortium |
| FOUNDING PROGRAM | H2020 |